Home > Products > Air Flow Calibrators Adapters > Afc Complete Series Digital Display With Selectable Options

AFC-COMPLETE-Series: Digital Display with Selectable Options

HI-Q's AFC-COMPLETE, digital air-flow calibrator, is "the field technicians first choice" when it comes to air-flow calibration. This microprocessor based air-flow calibrator incorporates the use of a differential pressure sensor, a precision thermistor, an absolute barometric pressure sensor, and a precision machined venturi tube, to measure and calculate instantaneous flow rate in either CFM*, LPM*, or CMM*, air temperature in either ºF or ºC, and Barometric Pressure in either Inches of Hg or mmHg. * "STP" may be selected for volumetric unit correction to Standard Temperature & Pressure.  *** Please Note that the AFC-COMPLETE-XX has been replaced by Model D-AFC-XX ***

  • Correction for Standard Temperature & Pressure at the Flip of a Switch Select "STP" or "Actual"
  • Temperature Measurement of "In-Line" Air Flow Display either F or C
  • Barometric Pressure Measurement Display Either Inch Hg or mm Hg
  • Durable Rotary Power/Unit Selection Switch Positive Option Selection
  • Large Digital Display 1/2" LCD
  • Battery Operated Unit May Be Run On Either AC or DC Power 

Specifications

Parameter Display Range Resolution
Barometer 0-31 in Hg;
0-770 mm Hg
0.01 in Hg
1 mm Hg
Air Flow See "Calibration Range" under "Ordering Information" 0.001CFM
Temperature 14 ºF to 140 ºF
-10 ºC to 60 ºC
0.04 ºF
0.02 ºC
Measurement Accuracy Flow Rate: ± 2.0% of Full Scale
Temperature: ±0.9ºF (± 0.5 ºC)
Barometric Pressure: ±1.5 mmHg @25 ºC
Flow Display Cubic Feet per Minute; Liters per Minute; Cubic Meters per Minute

Panel SwitchesRotary
(6 positions)(Slide = STP)
 

OFF, SCFM, SLPM, SCMM, TEMP (ºC), BARO (mm Hg)
(STP = Reading corrected to Standard Temperature & Pressure)
Panel SwitchesRotary
(6 positions)(Slide = ACTUAL)
OFF, ACFM, ALPM, ACMM, TEMP (ºF), BARO (in Hg)
(ACTUAL, "A" = Actual or Apparent flow reading)
Power Requirements Voltage, wall adapter: 12 VoltsCurrent: 80 milliampsBatteries:
Six "c" cells, 1.2 V NiCd
Overall Dimensions 12"W X 6"D X 11"H (Includes handle & inlet/outlet fittings)
Weight 8 pounds 4 ounces; (3.74 kg)
Storage Temperature -4ºF to 158ºF (-20°C to + 70°C) 
Operating Temperature 14ºF to 140ºF (-10°C to + 60°C)
Calibration Factory calibration is recommended once per year.
Note: Serial Port Calibration Module available under options.

Ordering Information

Select the desired flow range: Calibration Range Meter Connection
Model Number CFM LPM CMM
AFC-COMPLETE-1 0.1 to 1 3 to 30 0.003 to 0.030 3/8" Male QD
AFC-COMPLETE-2 0.2 to 2 6 to 56 0.006 to 0.056 3/8" Male QD
AFC-COMPLETE-3 0.3 to 3 9 to 85 0.009 to 0.085 3/8" Male QD
AFC-COMPLETE-4 0.4 to 4 11 to 113 0.011 to 0.113 3/8" Male QD
AFC-COMPLETE-5 0.5 to 5 14 to 142 0.014 to 0.142 3/8" Male QD
AFC-COMPLETE-10 1 to 10 28 to 280 0.028 to 0.280 3/8" Male QD
AFC-COMPLETE-15 2 to 15 56 to 425 0.056 to 0.425 1½" MSTP
AFC-COMPLETE-30 5 to 30 140 to 850 0.140 to 0.850 1½" MSTP
AFC-COMPLETE-40 5 to 40 140 to 1130 0.140 to 1.130 1½" MSTP
AFC-COMPLETE-50 5 to 50 140 to 1400 0.140 to 1.400 1½" MSTP
AFC-COMPLETE-70 10 to 70 280 to 1980 0.280 to 1.980 1½" MSTP
AFC-COMPLETE-XX
Replace "-XX" with uppermost flow rate.
Dependent upon flow range Dependent upon flow range Dependent upon flow range Dependent upon flow range

Note: For 230 Volt AC applications, add "/230" after model number.

Ordering Options

COMPLETE-CALMOD

"In-House" Calibration Module for AFC-COMPLETE Series Air Flow Calibrators. Connects to Serial Port on rear of unit.

COMPLETE-BATTERY

Spare Battery Pack for AFC-COMPLETE Series Air Flow Calibrators.